The DART 2.0: FGV-XY series gages from Larson Systems combine the extensive features of the DFS series with the rugged, ergonomic and compact design of the original DART series to provide a complete force testing instrument for a variety of industries. With free software and toriemon USB, the FGV-XY provides a total package for force testing up to 200lbs.
The ergonomically designed DART 2.0 provides an easy to use force gage for incoming quality inspection, finished goods testing, R&D and other force-related applications. Onboard statistic calculations provide fast access to maximum, minimum, average and standard deviation data.
The USB port and 1,000 point memory allows for data to be easily exported in real time or once testing is complete. In addition, the RS-232C and analog outputs allow the DART 2.0 series maximum flexibility.
Applications where the DART 2.0 series excels: Wallboard, wire pull, wire crimp, trigger pull, brake pedal, aerospace ceramic, spring, food ripeness, corrugated paper strength, blister pack medical component, muscle strength and ergonomic testing.
| Features | Benefits |
|---|---|
| 1000 Hz Sample Rate | Faster sampling to provide more accurate results |
| High Accuracy (0.2%) | Ideal for strict tolerances |
| USB/RS-232/Analog Outputs | 3 different outputs for data collection provides maximum flexibility |
| Reversible Keypad and Display | Use as a handheld gage or mount upside down on a stand No upside down reading required! |
| Variable display rate | Slow display update rate for operator ease |
| 0.5 Lb to 200 Lb Capacities | Various capacities allow testing of different applications without forfeiting accuracy |
| 200% Overload Capacity | Higher overload protection results in less repairs - Cheap to keep |
| Rugged Die Cast Aluminum Construction | Exceptional durability |
| Selectable Auto-Shut Off | Prolongs battery life, while allowing constant use when powered via AC adapter |
| Small Size | Ergonomic design provides comfort for hand held testing |
| On Board Statistics | Obtain immediate results and analysis of your tests |
| Expanded Memory | Store data points for later review and transfer to PC/Laptop |
| Built In Hi-Lo Comparator | Perform quick and easy ‘Go-No Go’ Testing |
| Model | FGV-0.5XY | FGV-1XY | FGV-2XY | FGV-5XY | FGV-10XY | FGV-20XY | FGV-50XY | FGV-100XY | FGV-200XY | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Capacity | 0.5 lb 2 N 200 g 8 oz |
1 lb 5 N 500 g 16 oz |
2 lb 10 N 1,000 g 32 oz |
5 lb 20 N 2 kg |
10 lb 50 N 5 kg |
20 lb 100 N 10 kg |
50 lb 200 N 20 kg |
100 lb 500 N 50 kg |
200 lb 1,000 N 100 kg |
|||
| Resolution | 0.001 lb 0.001 N 0.01 oz 0.1 g |
0.001 lb 0.01 N 0.01 oz 1 g |
0.001 lb 0.01 N 0.001 kg |
0.01 lb 0.01 N 0.001 kg |
0.01 lb 0.1 N 0.01 kg |
0.1 lb 0.1 N 0.01 kg |
0.1 lb 1 N 0.1 kg |
|||||
| Units of Measure | lb / oz / g / N | lb / kg / N | ||||||||||
| Measurement Modes | Standard, Compression Peak, Tension Peak | |||||||||||
| Display Rate | User Selectable: 1, 2, 3, 5, 10, 20 times per second | |||||||||||
| Sampling Rate | 1000 times per second | |||||||||||
| Accuracy | ± 0.2% F.S. | |||||||||||
| Influence of temperature | Gain : ± 0.01 LOAD / Zero : ±0.01 / R.C. / Drift of zero point can be cancelled with tracking function. | |||||||||||
| Display | Main display: 4-digits 12mm high, Reversible | |||||||||||
| Units display: 3-digits 7mm high | ||||||||||||
| Other display: "LO BAT"(Low Battery Voltage), "BAT"(Battery Charging), "OVR"(Overload), "Peak"(Peak Hold) | ||||||||||||
| Overload | 200% of Full Scale | 150% F.S. | ||||||||||
| Tracking | User Selectable (ON/OFF) | |||||||||||
| Output | USB | Allows communication between FGV-XY and Windows PC Software with ToriemonUSB via USB cable(included). | ||||||||||
| RS-232C | Allows communication between the FGV-XY and RS-232C devices. RS-232C cable is an optional accessory. | |||||||||||
| Analog | ± 1V, Accuracy is ± 50mV through a 12 bit D/A. ZERO affects this output, and is updated at 1000 times/second. Load is >10k Ω | |||||||||||
| Overload/ Comparator | Open-collector output (Max DC30V/5mA). | |||||||||||
| Power | Rechargeable Nickel hydride battery or AC adapter/charger. Usable while charging. Operating Time: Approximately 8 hours after a full charge. Charging Time: 16 Hours Max. |
|||||||||||
| Auto Power Off | Default is 10 minutes. Can be disabled. Automatically disabled when connected to AC adapter. | |||||||||||
| Memory function | Continuous memory: 1000 data points, Single memory: 100 data points, Standard memory: 50 data points Statistics functions (max, minimum, peak, average, standard deviation) | |||||||||||
| Comparator function | User Selectable: HI and LO | |||||||||||
| Temperature range | 0 ~ 40° | C (-32 ~ 104° F) | ||||||||||
| Humidity range | 35 ~ 85% RH | |||||||||||
| Dimensions | 147mm | (L)x 75mm (W) x 38mm(H) | ||||||||||
| Weight | Approx. 450g | Approx. 500g | ||||||||||
| Accessories (Included) |
AC adapter/charger, carrying case, hook, chisel, flat head, notched head, hanger, cone head, thread Adapter, | |||||||||||
| extension rod, USB cable | ||||||||||||
| Application software | Application software (USB version): Available at www.shimpoinst.com/software for free. | |||||||||||

| Ordering Information | ||
|---|---|---|
| Part Number | Description | Price |
| Force Gage | ||
| FGV-0.5XY - 0.5 lb DART 2.0 Force Gage | Quote | |
| FGV-1XY - 1 lb DART 2.0 Force Gage | Quote | |
| FGV-2XY - 2 lb DART 2.0 Force Gage | Quote | |
| FGV-5XY - 5 lb DART 2.0 Force Gage | Quote | |
| FGV-10XY - 10 lb DART 2.0 Force Gage | Quote | |
| FGV-20XY - 20 lb DART 2.0 Force Gage | Quote | |
| FGV-50XY - 50 lb DART 2.0 Force Gage | Quote | |
| FGV-100XY - 100 lb DART 2.0 Force Gage | Quote | |
| FGV-200XY - 200 lb DART 2.0 Force Gage | Quote | |
| Factory Options | ||
| NIST Traceable Dead Weight Calibration for FGV-XY Series | Quote | |